看过本文的还看了

相关文献

该作者的其他文献

文献详情 >The design of a cervical vertebra t... 收藏
The design of a cervical vertebra titanium plate-interbody fusion cage

The design of a cervical vertebra titanium plate-interbody fusion cage

作     者:Chuanyi Bai Kunzheng Wang Meigang Wei 

作者机构:Department of Orthopedics The Second Hospital of Xi'an Jiaotong University Xi'an 710004China 

出 版 物:《Journal of Nanjing Medical University》 (南京医科大学学报(英文版))

年 卷 期:2006年第20卷第2期

页      码:120-123页

摘      要:Objective: To study the biomechanical feature of a newly designed cervical vertebra internal fixation device and its clinical applications. Methods: Some functional spinal units were fixed respectively with titanium plate, fusion cage and new device designed by ourselves, then a controlled biomechanical study including flexion, extension, torsion and lateral bending was performed and the results were analyzed. Results: As to the mechanical performance, fusion cage showed poor performance in extension test and so did the titanium plate in the distortion test. However, the new device showed good performance in every test. Conclusion: Both simple titanium plate fixation and simple fusion cage fixation have biomechanical defaults, but they are complementary. The titanium plate-interbody fusion cage avoids the defaults and has specific advantages.

主 题 词:cervical vertebra internal fixation device biomechanics 

学科分类:10[医学] 

馆 藏 号:203118743...

读者评论 与其他读者分享你的观点

用户名:未登录
我的评分