文献详情 >A ballistocardiogram measurement sy... 收藏
A ballistocardiogram measurement system for home monitoring:design,performance,and evaluation

A ballistocardiogram measurement system for home monitoring:design,performance,and evaluation

作     者:Xinrong Cao Jie Wang Xianwen Zhang Jintian Tang 

作者机构:Institute of Medical Physics and EngineeringTsinghua University 

基  金:supported by the National Natural Science Foundation of C@hina (810701750) 

出 版 物:《Chinese Science Bulletin》 (Chin. Sci. Bull.)

年 卷 期:2014年第59卷第23期

页      码:2909-2917页

摘      要:We present a ballistocardiogram(BCG)measurement system for obtaining the BCG *** performance of the BCG measurement system was tested,including the dynamic response frequency of the platform(32 Hz with a 75 kg load),and the signal to noise ratio of the system(minimum 23 dB).Factors that could influence BCG recording during measurement were investigated by collecting data from 48 healthy subjects(28 males and 20females).The basic features of the BCG signal,such as the amplitude and time interval,were extracted,and differences by gender were analyzed.A correlation between the IJ amplitude(amplitude difference between I-and J-peak)and the respiration phase was found(R2=0.49).The BCG signals for both sitting(SiBCG)and standing(StBCG)postures were obtained using this *** showed that the StBCG amplitudes were larger than those of SiBCG but there was no significant difference in the time *** robustness of our system was verified,and its potential for non-invasive cardiac activity monitoring was demonstrated.

主 题 词:冲击测量系统 性能评价 信号传送 设计 监控 家庭 时间间隔 卡介苗 

学科分类:0711[理学-心理学类] 07[理学] 08[工学] 080401[080401] 0804[工学-材料学] 080402[080402] 

核心收录:

D O I:10.1007/s11434-014-0462-8

馆 藏 号:203141197...