文献详情 >Neutron collimator design of neutro... 收藏
Neutron collimator design of neutron radiography based on the BNCT facility

Neutron collimator design of neutron radiography based on the BNCT facility

作     者:杨晓鹏 俞伯祥 李义国 彭旦 鲁谨 张高龙 赵航 章爱武 李春阳 刘万金 胡涛 吕军光 

作者机构:School of Physics and Nuclear Energy Engineering Beihang University State Key Laboratory of Particle Detection and Electronics Institute of High Energy Physics Chinese Academy of Sciences China Institute of Atomic Energy University of South China 

基  金:Supported by National Natural Science Foundation of China(11105160) 

出 版 物:《Chinese Physics C》 (中国物理C(英文版))

年 卷 期:2014年第38卷第2期

页      码:97-100页

摘      要:For the research of CCD neutron radiography, a neutron collimator was designed based on the exit of thermal neutron of the Boron Neutron Capture Therapy(BNCT) reactor. Based on the Geant4 simulations, the preliminary choice of the size of the collimator was determined. The materials were selected according to the literature data. Then, a collimator was constructed and tested on site. The results of experiment and simulation show that the thermal neutron flux at the end of the neutron collimator is greater than 1.0×10^6n/cm^2/s, the maximum collimation ratio(L/D) is 58, the Cd-ratio(Mn) is 160 and the diameter of collimator end is 10 cm. This neutron collimator is considered to be applicable for neutron radiography.

主 题 词:neutron radiography ICCD camera thermal neutron collimator thermal neuron flux collimation ratio(L/D 

学科分类:0709[理学-地质学] 08[工学] 0708[理学-地球物理学类] 0804[工学-材料学] 0827[工学-食品科学与工程类] 082701[082701] 0703[理学-化学类] 0704[理学-天文学类] 0702[理学-物理学类] 

核心收录:

D O I:10.1088/1674-1137/38/2/028201

馆 藏 号:203227480...