Design, Realization and Implementation of a Novel Phantom for Quality Control of Micro-SPECT/CT and Micro-PET/CT Systems

Design, Realization and Implementation of a Novel Phantom for Quality Control of Micro-SPECT/CT and Micro-PET/CT Systems

作     者:Hatem Besbes Radhia Besbes Krid Bassel Solaiman 

作者机构:Laboratory of Biophysics and Medical Technology Faculty of Medicine Tunis El Manar University Tunis Tunisia Department of English Languages Institute of Gabes Gabes University Gabes Tunisia Department of Imagery and Information Processing Telecom Bretagne Brest France 

出 版 物:《Journal of Biomedical Science and Engineering》 (生物医学工程(英文))

年 卷 期:2015年第8卷第1期

页      码:46-55页

摘      要:The proposed phantom is designed for the quality control of micro-SPECT/CT and micro-PET/CT systems. However, it is an assembly of six patterns stored in a cylindrical box enabling to control both micro-SPECT unit in terms of uniformity, spatial linearity and spatial resolution and micro-CT unit in terms of uniformity, spatial linearity, spatial resolution, diffusion rate, low contrast detectability, Hounsfield unit linearity and slice thickness. The construction material is Plexiglas. As for the implementation, it was made on a micro-SPECT/CT machine of the type “speCZT eXplore CT 120”. Compared to the NEMA NU 4-2008 image quality phantom, this phantom offers micro-CT quality control and is more efficient in control of spatial resolution for micro-SPECT and micro- PET systems.

主 题 词:Phantom Quality Control Micro-SPECT/CT Micro-PET/CT 

学科分类:1002[医学-临床医学类] 100214[100214] 10[医学] 

D O I:10.4236/jbise.2015.81005

馆 藏 号:203458100...